Device Description

<< Click to Display Table of Contents >>

Navigation:  Hardware Overview > LOG Dataloggers > RMS-LOG-T30-868/915 >

Device Description

Functions

Description

Temperature Measurement

The measurement data are sent to the RMS WEB software at the set interval.

Logging function

oEvery measurement value and the time stamp is stored in the device memory.

oIn case of data interruption, the data can be read out via the RMS WEB software automatically afterwards.

Audit Trail

oThe device stores changes of the device settings.

oThese information are supplied to the RMS WEB Service.

Battery powered operation

If the external power supply fails, the device runs in battery mode. Measurements are still carried out in the specified interval, the data is stored in the memory and sent out to the RMS WEB service.

Firmware Update

The firmware update can be done by the RMS WEB Service.

 

 

Power Variants

Option 1: 24 V, via terminal (see "Accessories")

Option 2: Battery, see RMS-BAT RMS-BAT Rotronic Website

 

Power supply to be used: 24 VDC ±10 %, 4 W nominal, > 15 W Limited Power Source

 

 

Logging function

For each temperature measurement, the measured values are stored in the memory with the associated time stamp. At a measuring interval of one minute, data can be stored for one month, which corresponds to approx. 44,000 measurements. If the memory is full, the oldest values are overwritten.

 

 

Indicator

The device has a button and a multicolored LED for operation and display of the operating status. The button is used when the device is put into operation or to switch off the logging process in battery mode. The LED indicates the device status and whether a valid measurement was performed in the current measurement interval and the data could be sent to the monitoring system.

 

 

Push-button

The device is controlled and configured via a server application. When a device is integrated into a system, the user must confirm he is the owner of the respective device by pressing on a key.

 

Whether a device is integrated in the system or is primarily stored on the server. However, the device retains this information each time it communicates with the server and maintains the state in the event of a reboot or loss of connection. To remove the device from the system, it must be deleted from the server.

 

The push-button can be used to unpair as well.

 

 

RTCC (Real Time Clock Calendar)

The device is equipped with a real time clock. The time is continuously synchronized if there is a connection to the server.

 

 

Battery


Description

Article

RMS-BAT (ER14505M, various manufacturers - see website: accessories)

Type

Li-SOCl2

Capacity

~2000 mAh

Voltage

3.6 V

Dimension

AA (H: 50.3 mm, D: 14.55 mm)

After removing the wall bracket, insert the batteries according to the label in the battery compartment.

oThe batteries are used to power the unit in the event of a failure of the external power supply. In battery operation, the functionality is limited and the device cannot be addressed via Ethernet.

oLithium AA batteries with 3.6 V are used.

oThe mounting position must be observed. The polarity is indicated on the battery and in the battery compartment.

 

Rotronic have tested and recommend the following batteries:

1.Fanso ER14505M,

2.Titus ER14505M,

3.Ramway ER14505M.

 

When using other batteries, Rotronic cannot guaranty that the logger will work correctly.