Customer adjustment

<< Click to Display Table of Contents >>

Navigation:  Hardware Overview > LOG Dataloggers > RMS-LOG-T30-868/915 >

Customer adjustment

General

The customer can compensate possible deviations of the temperature measurement by an own adjustment. The customer uses temperature values as reference points for his adjustment.

 

 

1-point adjustment

1-point adjustment corresponds to a offset over the whole measurement range.


Calculation

Offset temperature

TOffset        = TRef - TAct


TRef        = reference temperature


TAct        = actual measured temperature (live from the device or noted by the customer)

End temperature

TEnd        = T(R) + TOffset


T(R)        = temperature (already corrected)

 

 

2-point adjustment

For the 2-point adjustment, the customer can acquire two points.  


Calculation

Slope

  Slope

Offset

intercept


TRef1        = reference temperature 1


TRef2        = reference temperature 2


TAct1        = measured temperature 1 (live from the device or noted by the customer)


TAct2        = measured temperature 2 (live from the device or noted by the customer)

End temperature

TEnd        = m * T(R) + n


T(R)        = temperature (already corrected)