The customer can compensate possible deviations of the temperature measurement by an own adjustment. The customer uses temperature values as reference points for his adjustment.
1-point adjustment corresponds to a offset over the whole measurement range.
Calculation |
|
Offset temperature |
TOffset = TRef - TAct |
TRef = reference temperature |
|
TAct = actual measured temperature (live from the device or noted by the customer) |
|
End temperature |
TEnd = T(R) + TOffset |
T(R) = temperature (already corrected) |
For the 2-point adjustment, the customer can acquire two points.
Calculation |
|
Slope |
|
Offset |
|
TRef1 = reference temperature 1 |
|
TRef2 = reference temperature 2 |
|
TAct1 = measured temperature 1 (live from the device or noted by the customer) |
|
TAct2 = measured temperature 2 (live from the device or noted by the customer) |
|
End temperature |
TEnd = m * T(R) + n |
T(R) = temperature (already corrected) |