<< Click to Display Table of Contents >> Navigation: Hardware Overview > LOG Dataloggers > RMS-LOG-T30-L > Change Log/Release Notes |
FW Version 1.3 (14.01.2026)
ID |
Description |
|---|---|
RMS-10977 |
Improved device time synchronization - correction of interval
In case of a difference between device time and RMS server time, a time synchronization happens from the server to the device. This time difference are caused mostly by time correction of the RMS Server cyclical. This time synchronization works fine.
The bug was that the interval from measurement to measurement was not corrected after the described time synchronization. After a long time of a running system, day by day, month by month, time synchronization could cause that the moment of measurement "wandered away" until a measurement jumped over to the next interval. In worst case, there was a missing measurement few times a year. In application of 10s interval e. g. for differential pressure measurement, it could be observed in the RMS system.
This bug fix now adapts the interval after a time synchronzation that this does not happen anymore.
Note: By design, time changes of server >5secs are not compensated anymore. Such big time shifts will be inherited and can cause data gaps. |
FW Version 1.2 (04.10.2022)
Summary
This new firmware has been implemented due to a quartz that was used for the real time clock has been discontinued and has been replaced with another internal quartz. Rotronic took advantage of this firmware update to implement an improved watchdog feature.
ID |
Description |
|---|---|
11217-3170 |
Quartz replacement |
11217-3171 |
Improved watchdog feature |
FW Version 1.2 - Initial version
Updated 15.01.2026