<< Click to Display Table of Contents >> Navigation: Hardware Overview > LOG Dataloggers > RMS-LOG-L > Change Log/Release Notes |
FW Version 1.7 (14.01.2026)
ID |
Description |
|---|---|
RMS-10868A |
Timing Uart adapt for HCD-IC The communication between the logger and the connected digital probe was improved / stabilized for extreme climates. |
RMS-10977 |
Improved device time synchronization - correction of interval
In case of a difference between device time and RMS server time, a time synchronization happens from the server to the device. This time difference are caused mostly by time correction of the RMS Server cyclical. This time synchronization works fine.
The bug was that the interval from measurement to measurement was not corrected after the described time synchronization. After a long time of a running system, day by day, month by month, time synchronization could cause that the moment of measurement "wandered away" until a measurement jumped over to the next interval. In worst case, there was a missing measurement few times a year. In application of 10s interval e. g. for differential pressure measurement, it could be observed in the RMS system.
This bug fix now adapts the interval after a time synchronzation that this does not happen anymore.
Note: By design, time changes of server >5secs are not compensated anymore. Such big time shifts will be inherited and can cause data gaps. |
FW Version 1.6 (5th of October 2022)
Summery
This new firmware has been implemented due to a quartz that was used for the real time clock has been discontinued and has been replaced with another internal quartz. Rotronic took advantage of this firmware update to implement an improved watchdog feature.
ID |
Description |
|---|---|
11064-3182 |
Timing Uart adapt for HCD-IC The communication between the logger and the connected digital probe was improved / stabilized for extreme climates. |
11064-3183 |
Improved watchdog feature. |
11064-3097 |
When a data gap occurs, the user can push the try again button. Then the command to download the data from the logger is sent. Normally, if no data is available, "no data" will appear in the audit trail. With the RMS-LOG-L a "timeout" will appear instead of "no data".ers shows timeout instead of no data. |
11064-2549 |
The RMS-LOG-L will not accept incorrect IP configurations. |
11064-1121 |
With a key refresh, a new key is generated and the response is sent back with the new key. This interrupts the communication between the logger and the web service. |
11064-1088 |
The Peripheral Module Disable Bits are only masked on a byte-by-byte basis, but should be masked on a word-by-word basis. |
11064-991 |
The define PMD_ALL_OFF is only defined as a byte and not as a word. The high byte of the PMD register is therefore not set and the corresponding modules are not switched off. This could further reduce the sleep current. |
11064-851 |
The manual write protection is automatically switched off in the new version of RMS-CONFIG. The feature is obsolete and can be removed from the source code. |
11064-3121 |
Restore Default Properties in case of Reset the device and press the button. Several RMS-LAN-devices is this feature already implemented. |
FW Version 1.5
•Integration of follow sensors
▪HCD (available)
▪HCD-IC (coming soon)
▪PCD-S CO2 Sensor (Q2 2018)
▪CCD-S differential pressure (Q2 2018)
▪LDP
•For operation of sensors with increased power consumption, the current limit at the sensor interface has been increased. A maximum continuous current of 70mA can be supplied
•For the operation of sensor with long measuring times, the sensor can activate a permanent supply
•The communication via HTTP to the web service (from V1.21) is encrypted packet content. The key exchange takes place with ECDH procedure, which is encrypted according to AES-128 CBC.
•During the pairing new additional information about the connected sensor will be transmitted. The web service (from V1.2) can thus suggest suitable measuring points during the pairing.
Bugs fix
•For serial numbers above 2147483647, the number in the DHCP host name was displayed as a negative number.
FW Version 1.3
Summary
•The device now has a galvanically isolated supply. The voltage regulator originally used has been replaced as it has been discontinued.
•The following parameters can be read via the Modbus TCP standard commands (Function 4, Read Input Register):
▪Serial number of the device
▪Serial number of the connected sensor
▪Measured values 1 and 2 of the connected sensor
•When logging in to the DHCP server, the host name is specified in the form "RMS-LOG-L-00000000".
•Improve low-battery detection with a comparator
•If the network settings are written via the input mask in the device search for another device, the device restarts.
•When changing the date from the 30th to the 31st in the months of October and December, the 30th was changed directly to the 1st.
ID |
Description |
|---|---|
11064-326 |
RTCC, month change in Oct, Dec on 30th, instead of 31st. |
11064-325 |
Discovery, reset when writing network settings on another device |
FW Version 1.2 - Initial version
Updated 15.01.2026