Change Log/Release Notes

<< Click to Display Table of Contents >>

Navigation:  Hardware Overview > Input and Output Devices > RMS-ADC-L-R >

Change Log/Release Notes

Version Overview

Please find below the meaning of the version.

1: Major version number.

X.1: Minor version number.

 

Rotronic Tracking System

Rotronic use a tracking system to monitor bugs, features, improvements and change requests. This tracking system is the basis for any software, hardware and firmware User Requirement Specification based upon the GAMP©5 validation model. Each point will be recognized with an ID (eg. RMS-XXX). A description and details are attributed to each point within the tracking system. All changes documented within the RMS Release Notes will have an ID number and a description.

 

1.Version V1.3.

 

a.Release Details

The main target of this firmware version has been to correct an issue with the device time stamp.

 

b.Bug Fixes

ID

Description

RMS-2426A

The time synchronization between the device time and server time, when the server time changed between 1 and 5 sec caused a change of the time stamp of the measured values and data gaps appeared.

 

2.Version V1.2

 

WARNING: When using the firmware V1.1, an upgrade or downgrade of the firmware is not possible. The device must be returned to Rotronic AG to get a firmware update.

 

a.Release Details

This new firmware has been implemented due to a quartz that was used for the real time clock has been discontinued and has been replaced with another internal quartz. Rotronic took advantage of this firmware update to implement an improved watchdog feature.

 

Important: due to the quartz update, it is no longer possible to downgrade the firmware on new devices.

 

b.New Features

ID

Description

11122-3172

Quartz replacement.

11122-3173

Improved watchdog feature.

 

3.Version V1.1

 

WARNING: When using the firmware V1.1, an upgrade or downgrade of the firmware is not possible. The device must be returned to Rotronic AG to get a firmware update.

 

a.Release Details

The main target of this firmware version has been to correct existing bugs that can occur when an RMS-ADC-L-R gets an overvoltage spike.

 

Important: Please use RMS ≥V2.1 and update the device properties.

 

b.New Features

ID

Description

11122-2876

Audit Trail and Status Log implementation.

 

c.Bug Fixes

ID

Description

11122-2602

Overvoltage protection at the analogue inputs causes a measurement error. The protection was ensured with PCB improvement.

 

4.Version V1.0

Initial release.

 

Updated: 15.04.2024.