Carry out an Adjustment

<< Click to Display Table of Contents >>

Navigation:  Software Overview > RMS-CONFIG Software > Software Functions > Device > Device Settings > Adjust >

Carry out an Adjustment

Within RMS-CONFIG it is important to understand how the adjustment function works.

 

oA 1 point adjustment is an offset.

oA 2 point adjustment is an offset and a slope.

oA >2 point adjustment is an offset, a slope and smoothing.

 

When adjusting temperature, a maximum of 2 points can be adjusted.

 

Adjustment process.

 

1.Delete all points acquired within the device.

2.After the defined time for the calibration, add a reference value and click on acquire.

3.Should a 1 point adjustment be required, press on adjust.

4.Should a 2 point adjustment be required, after the defined time for the calibration, add a reference value and click on acquire.

5.Should a 2 point adjustment be required, press on adjust.

6.Should a >2 point adjustment be required, after the defined time for the calibration, add a reference value and click on acquire. Repeat this step for the desired amount of adjustments.

7.When all points have been acquired, press on adjust.