Within RMS-CONFIG it is important to understand how the adjustment function works.
oA 1 point adjustment is an offset.
oA 2 point adjustment is an offset and a slope.
oA >2 point adjustment is an offset, a slope and smoothing.
When adjusting temperature, a maximum of 2 points can be adjusted.
Adjustment process.
1.Delete all points acquired within the device.
2.After the defined time for the calibration, add a reference value and click on acquire.
3.Should a 1 point adjustment be required, press on adjust.
4.Should a 2 point adjustment be required, after the defined time for the calibration, add a reference value and click on acquire.
5.Should a 2 point adjustment be required, press on adjust.
6.Should a >2 point adjustment be required, after the defined time for the calibration, add a reference value and click on acquire. Repeat this step for the desired amount of adjustments.
7.When all points have been acquired, press on adjust.