<< Click to Display Table of Contents >> Navigation: Hardware Overview > WiFi Dataloggers > RMS-LOG-W-T10-D > Accuracy |
The accuracy of a RMS-LOG-W-T10-D is effected by
•Analog-Digital-Convertion (ADC),
•Used sensor type
•Resistance-Temperature-Calculation (based on sensor table or parameters)
•Long term drift of the sensor
•and other influences as e. g. adjustment and its related reference accuracy
Topic |
Description |
---|---|
Accuracy ADC |
<0.1°C (range of -70...+165°C for T10-0002) |
Ambient temperature influence |
<0.1°C (range of -30...+50°C ambient temperature of the logger) |
Please see chapter Temperature NTC
To calculate the a measurement uncertainty for an individual application, all known accuracies/failure influences might be known and calculated by Pythagorean Addition of the known accuracies/failure influences.
Update: 27.05.2025